Internal Internal

Do you already have an account? Log in

X Đóng lại
Search

Applications

Steam Autoclave is designed and manufactured for the purpose of sterilizing:
- Medical instruments, surgical instruments, medical supplies, medical materials, etc…
- Protective clothes, caps, gloves, mask, etc for medical doctors, medical staffs and patient, etc…
- Instrument, materials for test, experiment in laboratory…
- Liquid in medicine, test, experimental laboratory… in both wrapped and unwrapped types.

Main features

- Operating by a high precision multifunctional microprocessor controller, user interface via LCD display and keypad.
- Automatic operations from start to end of sterilizing cycle: water supply – pre vacuum – heating – sterilizing – ventilation – post drying – air balance – finish (end cycle).
- Sterilant: Saturated steam at high temperature.
- Equipped with PT 100 sensor to monitor temperature of the sterilization chamber.
- Display on LCD display for: Name of phases of the sterilization cycle. Chart of phases of the sterilization cycle. Pre-set temperature and actually operating temprature. Pre-set time (sterilization, dry) and remaining time.
- Actual sterilization pressure in the process of operation is indicated in pressure gauge.
- Alarms by signal lights and display alarm message on LCD display.
Equipped with door locking system and protection function for temperature, pressure, sensor, control signal, water level, electric current, etc…

Technical data

Model
 
HA-20L
HA-20DR
HA-25L
HA-25DR
HA-30L
HA-30DR
HA-45L
HA-45DR
HA55L
HA-55DR
HA-60L
HA-60DR
HA-75L
HA-75DR
HA-100L
HA-100DR
Type
Horizontal, Front Door
Floor Standing, Upper Door (Top Lid).
Chamber
Horizontal Cylinder
Vertical Cylinder.
Volume
20 litre
25 litre
30 litre
45 litre
55 litre
60 litre
75 litre
100 litre
Temp. & Time Controller
Fully Automatic Microprocessor Controller.
Programs
3 Pre-set Sterilization Programs (for wrapped articles, unwrapped articles, liquids) but resettable
5 Pre-set Sterilization Programs (for wrapped articles, unwrapped articles, liquids) and
1 Programe reserve for the user himself set-up.
Terilization Temp. Setting Range
105°C ÷ 135°C
Sterilization Pressure Range
0.5 ÷ 2.3 kg/cm² (equivalent to 0.49 bar – 2.256 bar).
Sterilization Time Setting Range
1 ÷ 180 minutes. Setting scale: 1 minute.
Drying Time Setting Range
0 ÷ 180 minutes. Setting scale: 1 minute.
Drying Method
Electric.
Electric (Model: HA-xxxL)
Vacuum Pump (Model: HA-xxxDR)

Catalogue/Brochure

** Please do not hesitate to contact us for more informations